Wet Film Comb is Used for checking the coating thickness of wet paint,enamel lacquer, adhensive , architectural coating, color coating, varnish, chemical industry, or other coating that has been applied to a smooth surface.

Standards of Wet Film Comb
GB/T1764-79 GB/T134522-92 ISO2808-74
Range
25~3000μm / 1~118mil
Values
25, 50, 75, 100, 125, 150, 175, 200, 225, 250, 275, 300, 350, 400, 450, 500, 550, 600, 650, 700, 750, 800, 850, 900, 1000, 1100, 1200, 1400, 1600, 1800, 2000, 2200, 2400, 2600, 2800, 3000µm
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 14 ,16, 18, 20, 22, 24, 26, 28, 30, 31, 33, 35, 39, 43, 47, 55, 63, 71, 79, 87, 94, 102, 110, 118mil

1. Select comb according to approximate thickness.
2. Hold comb perpendicular to coating to be measured.
3. Press into coating until the comb stands square.
4. Wet film thickness is between the highest step coated
and the next (higher) step remains uncoated.

GB/T1764-79 GB/T134522-92 ISO2808-74

25~3000μm / 1~118mil

Standards
GB/T1764-79 GB/T134522-92 ISO2808-74
Range
25~2032μm / 1~80mil
Values
25, 51, 75, 102, 127, 152, 178, 203, 229, 254, 279, 305, 356, 406, 457, 508, 559, 610, 660, 711, 762, 889, 1016, 1143, 1270, 1397, 1524, 1651, 1778, 2032µm